Application Note | ElastoSens™ BIO
How to measure ex-vivo tissue membranes with ElastoSens™ Bio
by Dimitria Camasão,
Application Scientists, Rheolution Inc.
Summary
- The mechanical characterization of biological soft membranes such as skin, pericardium, or hydrogel-based samples is challenging and often results in sample damage.
- The Membrane Sample Holder of the ElastoSens™ Bio has been designed for facilitating the loading and the handling of membrane samples allowing proper mechanical testing.
- In this study, the elasticity of bovine pericardium membranes were precisely measured using the ElastoSens™ Bio.
Mechanical characterization of ex-vivo tissue membranes
Introduction
Membrane-type samples are widely present in the life sciences field and are often difficult to characterize mechanically. Examples include native and engineered tissues such as pericardium, skin, placenta, epithelial and connective membranes as well as dressings and patches for wound care, suture or transdermal drug delivery. Due to their soft nature, their manipulation and introduction into analytical instruments often result in sample damage.
In this light, the Membrane Sample Holder of the ElastoSens™ Bio was designed to test the mechanical properties of synthetic and biological soft membranes. The Membrane Sample Holder is completed with a preparation kit that facilitates the loading and the handling of the membrane to avoid its damage and ensure a proper introduction into the testing instrument. In addition, all components are autoclavable and the same sample can be tested over time and as needed due to the non-destructive nature of the ElastoSens™ Bio. Samples contained into the Membrane Sample Holder can be stored in 12-well plates between tests.
Evaluating pericardium membranes: preparation and mechanical testing
Materials and Methods
Human and animal pericardium are explored in the biomedical field as grafts for the repair of damaged tissues or as matrices for tissue engineering purposes. Bovine pericardium membranes without (type I) and with (type II) a crosslinking treatment were cut into circles (diameter of approximately 2 cm) and introduced in PBS for 1 hour before measurements at room temperature. Sample thickness was measured using laser profilometry with the sample over a flat surface (Figure 1). The type I membrane had a measured thickness of 0.46 ± 0.02 mm, while the type II membrane displayed a slightly thicker dimension of 0.61 ± 0.04. Samples were then clamped into the stainless steel membrane holding grip of the membrane sample holder (Figure 2). The tests were immediately performed at room temperature in the ElastoSens™ Bio.
Figure 1: Measurement of sample thickness using a laser.
Figure 2: Pericardium membranes mounted in the membrane holding grip of the membrane sample holder.
Shear storage modulus (G’) was obtained for each membrane type (n=3). Average results are expressed as mean ± standard deviation. Comparisons among groups were evaluated by Welch’s t-test. Significance was retained when p < 0.05.
High-precision mechanical testing of pericardium membranes
Results and Discussion
Figure 3 displays the average G’ (kPa) for each sample type (n=3). Pericardium membranes that received crosslinking treatment showed a G’ three times higher than the membranes without the treatment. It is important to note that a statistical difference was obtained (p=0.008) even with the inherent intra variability of animal tissues and the small number of samples (n=3), reinforcing the high repeatability and sensitivity of the results.
Figure 3: Average G’ (kPa) of bovine pericardium membranes without (type I) and with (type II) a crosslinking treatment (n=3). ** p=0.008
Conclusions and Perspectives
Biological membrane tissues were successfully loaded and tested ex vivo into the membrane sample holder of the ElastoSens™ Bio. This study illustrates the high repeatability and sensitivity of the instrument when using the membrane sample holder to measure this type of samples.
Impact & Use
- Biological and synthetic membrane samples can be easily and precisely measured with the membrane sample holder of the ElastoSens™ Bio.
- After testing, membrane samples can be stored in 12-well plates and measured over time to study their mechanical stability under controlled conditions.
- The effect of different formulations, manufacturing techniques and treatments on membrane samples can be easily studied using the viscoelastic properties obtained with the ElastoSens™ Bio.
- The mechanical properties of biological tissues and organs can effectively be tested ex vivo using the membrane sample holder of the ElastoSens™ Bio.
- The heterogeneity of membrane tissues in terms of mechanical properties can be mapped using the ElastoSens™ Bio.
- The viscoelastic properties obtained with the ElastoSens™ Bio is an important parameter to determine the variability of biological tissues (within the same organ, between different organs, and among species).
Discover how our technology non-destructively measures the viscoelastic properties of soft biomaterials and tissues using micro-volumes of samples
Related Posts
The peritoneal membrane is a thin, semi-permeable tissue whose mechanical behavior is closely tied to its structure, vascular network, and ability to withstand pressure-driven fluid movement. During normal function or during peritoneal dialysis, the membrane is continuously subjected to osmotic gradients, hydrostatic pressures, and shifts in tissue hydration—each of which reveals aspects of its mechanical performance.
The pericardium plays a central role in modulating how the heart fills, moves, and interacts with its surrounding structures. Its mechanical behavior—defined by properties such as elasticity, stiffness, anisotropy, and resistance to bending—helps maintain the heart’s shape, stabilize its position, and distribute load during every cardiac cycle.
The pancreas is a small but complex organ whose function depends not only on biochemical signaling but also on the physical environment of its tissue. Its mechanical properties—such as stiffness, elasticity, and viscoelasticity—are shaped by the structure of the parenchyma and the extracellular matrix.
The aorta is a highly specialized elastic artery designed to buffer the pulsatile output of the heart and maintain smooth, continuous blood flow throughout the body. Its wall is composed of elastin, collagen, smooth muscle cells, and connective tissue arranged in a layered structure that gives rise to complex mechanical behavior.
The brain’s ability to regulate cognitive, sensory, and motor functions depends not only on its complex network of neurons and signaling pathways but also on the unique mechanical behavior of its tissue. Brain matter is exceptionally soft, heterogeneous, and sensitive to deformation, which makes its mechanical properties central to both normal function and vulnerability to injury.
The spleen is a soft, vascular organ whose structure is closely linked to its roles in blood filtration, immune defense, and hematological balance. Its tissue mechanics—such as stiffness, elasticity, and viscoelasticity—are important markers of both normal physiology and pathological change. Alterations in these properties often mirror systemic conditions, making the spleen an informative window into overall health.