Application Note | ElastoSens™ Bio
Viscoelasticity characterization of coagulating plasma using ElastoSens™ Bio
by Dr. Dimitria Camasão
Senior Application Specialist, Rheolution Inc.
NOTICE
This study was a joint collaboration between Rheolution Inc. (QC, Canada) and PR. Showan Nazhat’s group (Department of Mining and Materials Engineering, McGill University, QC, Canada).
SUMMARY
— Clotting tests, such as prothrombin time (PT) and activated partial thromboplastin time (aPTT), are frequently used in clinics and research to identify coagulation disorders and to guide coagulant and anticoagulant drug development.
— ElastoSens™ Bio successfully measured the viscoelastic properties of coagulating animal plasma when mixed with PT and aPTT reagents.
— The activation of the extrinsic pathway (with PT) led to faster plasma coagulation and to a stiffer clot when compared with the activation of the intrinsic pathway (with aPTT).
INTRODUCTION
The human body has a natural mechanism to stop bleeding after an injury. Platelets migrate to the site of injury and start to form a soft blood clot. This activates other clotting factors in the bloodstream triggering a chain reaction to form a harder blood clot that will stay firmly in place.
Coagulation disorders can result in two opposite conditions: the blood clots easily even without an injury or takes too long to clot after an injury. The first condition may lead to abnormal clots flowing freely within blood vessels that could potentially block the blood flow and cause serious complications such as heart attack and stroke. The second condition can lead to serious blood loss after an injury or surgery.
Coagulation disorders of a patient are detected with blood or plasma clotting tests such as prothrombin time (PT) and activated partial thromboplastin time (aPTT). The results of this test are compared to reference values to help determine the coagulation state of the patient and guide an optimal treatment. In research, these tests can also be used for the development of coagulant and anticoagulant drugs and technologies.
In this application note, the ElastoSens™ Bio was used to investigate the coagulation kinetics of blood plasma mixed with commercial PT and aPTT reagents in terms of viscoelastic properties.
MATERIALS AND METHODS
Pacific Hemostasis® Thromboplastin-DS and Pacific Hemostasis® APTT-XL (Fisher Diagnostics®, Thermo Fisher Scientific Inc., ON, Canada) were used to induce the extrinsic and the intrinsic coagulation pathway of calf plasma (Cedarlane, ON, Canada), respectively. Sample preparation was performed according to the manufacturer guidelines. Briefly, plasma was thawed at 37 °C and reagents were prewarmed at the same temperature. Plasma and reagents were mixed in the ratio of 1:2 (2.5 mL of plasma and 5.0 mL of reagents), transferred into the sample holder of the ElastoSens™ Bio (Rheolution Inc., QC, Canada), and the test in the soft mode was immediately started at 37 °C for 300 s.
The shear storage modulus (G’) as function of time of each sample was analyzed for the determination of the following descriptors:
Average results are expressed as mean ± standard deviation (n = 2). Statistical analysis was performed using GraphPad Prism version 9 (GraphPad Prism Software, CA, USA). Unpaired t-test with Welch’s correction was applied.
RESULTS AND DISCUSSION
Figure 1 shows the shear storage modulus (G’) of animal plasma with PT (blue) and aPTT (orange) reagents as function of time. For both reagents, G’ started close to 0 indicating that the sample was still liquid in the beginning of the test. The G’ then gradually increased as a function of time. PT and aPTT trigger the extrinsic and the intrinsic pathway of coagulation, respectively. The extrinsic pathway is shorter, so it is expected that the coagulation happens faster when compared to the intrinsic pathway once activated separately in vitro [1]. Indeed, G’ began to increase before (i.e. clot started to form faster) for the plasma mixed with the PT reagent when compared to the aPTT reagent. All samples formed a clot by the end of the test (Figure 2).
Figure 1: Shear storage modulus (G’) as a function of time of animal plasma with PT (blue) and aPTT (orange) reagents.
Figure 2: Plasma clot at the end of the test.
Figure 3 shows that clotting initiation time was 28.2 ± 4.8 seconds for the PT against 62.9 ± 3.0 seconds for the aPTT (p<0.05). The maximum rate of clot formation was higher for the PT (4.3 ± 0.4 Pa/s) when compared to the aPTT (2.3 ± 0.1 Pa/s). The final clot was also different in terms of final stiffness. Average G’ in the plateau region of the curves were 162.3 ± 13.8 Pa for the PT reagent and 106.3 ± 2.2 Pa for the aPTT reagent. The clot stiffness is related to the components initially present in the sample and their interactions. Higher stiffness can suggest that a higher amount and/or stronger links among the plasma components (such as fibrinogen, thrombin, and clotting factors) happened.
Figure 3: Clotting initiation time, maximum rate of clot formation, and final clot stiffness of calf plasma with PT (blue) and aPTT (orange) reagents.
CONCLUSIONS
ElastoSens™ Bio successfully measured the viscoelastic properties of coagulating animal plasma when mixed with PT and aPTT reagents. Animal plasma started to coagulate faster, had a higher rate of formation, and resulted in a stiffer clot when mixed with the PT reagent when compared to the aPTT reagent.
PERSPECTIVES
— ElastoSens™ Bio is able to capture subtle changes in very soft samples such as coagulating plasma.
— The stiffness of the formed plasma clot can also provide valuable information about the extrinsic and intrinsic coagulation pathways as it is directly related to the components present in the sample and their interactions.
— ElastoSens™ Bio can be used in preclinical studies to investigate the effect of coagulant and anticoagulant drugs on plasma.
— The stability of the formed plasma clots can be monitored over time with the same sample using the ElastoSens™ Bio.
REFERENCES
[1] Chaudhry, R., Usama, S. M., & Babiker, H. M. (2018). Physiology, coagulation pathways.
INTRODUCTION
The human body has a natural mechanism to stop bleeding after an injury. Platelets migrate to the site of injury and start to form a soft blood clot. This activates other clotting factors in the bloodstream triggering a chain reaction to form a harder blood clot that will stay firmly in place.
Coagulation disorders can result in two opposite conditions: the blood clots easily even without an injury or takes too long to clot after an injury. The first condition may lead to abnormal clots flowing freely within blood vessels that could potentially block the blood flow and cause serious complications such as heart attack and stroke. The second condition can lead to serious blood loss after an injury or surgery.
Coagulation disorders of a patient are detected with blood or plasma clotting tests such as prothrombin time (PT) and activated partial thromboplastin time (aPTT). The results of this test are compared to reference values to help determine the coagulation state of the patient and guide an optimal treatment. In research, these tests can also be used for the development of coagulant and anticoagulant drugs and technologies.
In this application note, the ElastoSens™ Bio was used to investigate the coagulation kinetics of blood plasma mixed with commercial PT and aPTT reagents in terms of viscoelastic properties.
MATERIALS AND METHODS
Pacific Hemostasis® Thromboplastin-DS and Pacific Hemostasis® APTT-XL (Fisher Diagnostics®, Thermo Fisher Scientific Inc., ON, Canada) were used to induce the extrinsic and the intrinsic coagulation pathway of calf plasma (Cedarlane, ON, Canada), respectively. Sample preparation was performed according to the manufacturer guidelines. Briefly, plasma was thawed at 37 °C and reagents were prewarmed at the same temperature. Plasma and reagents were mixed in the ratio of 1:2 (2.5 mL of plasma and 5.0 mL of reagents), transferred into the sample holder of the ElastoSens™ Bio (Rheolution Inc., QC, Canada), and the test in the soft mode was immediately started at 37 °C for 300 s.
The shear storage modulus (G’) as function of time of each sample was analyzed for the determination of the following descriptors:
Average results are expressed as mean ± standard deviation (n = 2). Statistical analysis was performed using GraphPad Prism version 9 (GraphPad Prism Software, CA, USA). Unpaired t-test with Welch’s correction was applied.
RESULTS AND DISCUSSION
Figure 1 shows the shear storage modulus (G’) of animal plasma with PT (blue) and aPTT (orange) reagents as function of time. For both reagents, G’ started close to 0 indicating that the sample was still liquid in the beginning of the test. The G’ then gradually increased as a function of time. PT and aPTT trigger the extrinsic and the intrinsic pathway of coagulation, respectively. The extrinsic pathway is shorter, so it is expected that the coagulation happens faster when compared to the intrinsic pathway once activated separately in vitro [1]. Indeed, G’ began to increase before (i.e. clot started to form faster) for the plasma mixed with the PT reagent when compared to the aPTT reagent. All samples formed a clot by the end of the test (Figure 2).
Figure 1: Shear storage modulus (G’) as a function of time of animal plasma with PT (blue) and aPTT (orange) reagents.
Figure 2: Plasma clot at the end of the test.
Figure 3 shows that clotting initiation time was 28.2 ± 4.8 seconds for the PT against 62.9 ± 3.0 seconds for the aPTT (p<0.05). The maximum rate of clot formation was higher for the PT (4.3 ± 0.4 Pa/s) when compared to the aPTT (2.3 ± 0.1 Pa/s). The final clot was also different in terms of final stiffness. Average G’ in the plateau region of the curves were 162.3 ± 13.8 Pa for the PT reagent and 106.3 ± 2.2 Pa for the aPTT reagent. The clot stiffness is related to the components initially present in the sample and their interactions. Higher stiffness can suggest that a higher amount and/or stronger links among the plasma components (such as fibrinogen, thrombin, and clotting factors) happened.
Figure 3: Clotting initiation time, maximum rate of clot formation, and final clot stiffness of calf plasma with PT (blue) and aPTT (orange) reagents.
CONCLUSIONS
ElastoSens™ Bio successfully measured the viscoelastic properties of coagulating animal plasma when mixed with PT and aPTT reagents. Animal plasma started to coagulate faster, had a higher rate of formation, and resulted in a stiffer clot when mixed with the PT reagent when compared to the aPTT reagent.
PERSPECTIVES
— ElastoSens™ Bio is able to capture subtle changes in very soft samples such as coagulating plasma.
— The stiffness of the formed plasma clot can also provide valuable information about the extrinsic and intrinsic coagulation pathways as it is directly related to the components present in the sample and their interactions.
— ElastoSens™ Bio can be used in preclinical studies to investigate the effect of coagulant and anticoagulant drugs on plasma.
— The stability of the formed plasma clots can be monitored over time with the same sample using the ElastoSens™ Bio.
REFERENCES
[1] Chaudhry, R., Usama, S. M., & Babiker, H. M. (2018). Physiology, coagulation pathways.
Related Application Notes
We are your partners in viscoelasticity testing. That’s why our expert corner will be sharing with you, every 3 months, a curated selection of summarized scientific articles and original articles from our team:
- To make your life easier and save you time
- To keep you informed about what’s new in viscoelasticity testing
- To learn more about the various applications of biomaterials